  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Users
Manual
DPharp
Differential Pressure and
Pressure Transmitters
Manual Change No. 13-005
June 28, 2013
(1)
Please refer to this manual change for the handling cautions of ATEX certication if the products include
the option code /KF21 or /KS25.
IM No. Model & Option code
Applicable Part of
this Manual Change
IM 01C25T02-01E
EJXA with option code /KF21 or /KS25
( I )
IM 01C25R03-01E
(2)
Please use this manual change for the page listed below.
IM No. Page Contents
Applicable Part of
this Manual Change
IM 01C25T02-01E 6-9
6.4.6 Units That Can be Displayed on the
LCD by the Automatic Link Function
( II )
Vista de pagina 0
1 2 3 4 5 6 7 8 9

Indice de contenidos

Pagina 1 - Pressure Transmitters

User’sManualDPharpDifferential Pressure andPressure TransmittersManual Change No. 13-005June 28, 2013(1)Please refer to this manual change for the han

Pagina 2 -  FISCO Model

( I )ATEX Certication(1) Technical Dataa. ATEX Intrinsically Safe TypeCaution for ATEX Intrinsically safe type.Note 1. EJX series pressure transmi

Pagina 3 -  Entity Model

The criterion for such interconnection is that the voltage (Ui), the current (Ii) and the power (Pi), which intrinsically safe apparatus can receive,

Pagina 4 - * –15ºC when /HE is specied

Electrical Data• When combined with Trapezoidal output characteristic FISCO model IIC barrier [Supply/Output circuit (terminals + and –)] Ui = 17.5

Pagina 5 - Terminal

Supply unitThe supply unit must be certied by a Notied body as FISCO model and following trapezoidal or rectangular output characteristic is used.Uo

Pagina 6 - ISO M20 × 1.5 female

Note 4. Operation• Keep the “WARNING” label attached to the transmitter. WARNING: AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE OPENING. WHEN THE AMB

Pagina 7 - F0211r.ai

 Name plate Tag plate for flameproof type Tag plate for intrinsically safe type Tag plate for intrinsically safe type for EJX90AF0211r.aiDDAFTER

Pagina 8

INDEX Unit Display1138 mbar mbar1139 torr torr1140 atm atm1141 psi psi1142 psia psia 1143 psig psig 1144 g/cm2g/cm21145 kg/cm2kg/cm21146 inH2O inH2O

Pagina 9

INDEX Unit Display1502* Mm3/min Mm3/min1503* µm3/h um3/h1504* mm3/h mm3/h1505* km3/h km3/h1506* Mm3/h Mm3/h1507* µm3/d um3/d1508* mm3/d mm3/d1509* km3