Note 2. Electrical Data
Supply voltage: 42 V dc max.
Output signal: 4 to 20 mA
Note 3. Installation
All wiring shall comply with local installation
requirement.
The cable entry devices shall be of a certied
ameproof type, suitable for the conditions of
use.
Note 4. Operation
Keep the “WARNINGlabel attached to the
transmitter.
WARNING: AFTER DE-ENERGIZING,
DELAY 5 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP.≥65°C, USE
HEAT-RESISTING CABLES≥90°C.
Take care not to generate mechanical
sparking when accessing to the instrument
and peripheral devices in a hazardous
location.
Note 5. Maintenance and Repair
The instrument modication or part
replacement by other than an authorized
representative of Yokogawa Electric
Corporation is prohibited and will void KEMA
Flameproof Certication.
WARNING
To satisfy IP66 or IP67, apply waterproof glands
to the electrical connection port.
(2) Electrical Connection
A mark indicating the electrical connection type
is stamped near the electrical connection port.
These marks are as followed.
F0208r.ai
Location of the mark
Screw Size Marking
ISO M20 × 1.5 female
ANSI 1/2 NPT female
M
A or WN or
(3) Installation
WARNING
All wiring shall comply with local installation
requirements and the local electrical code.
There is no need for conduit seal in Division
1 and Division 2 hazardous locations
because this product is sealed at the factory.
(4) Operation
WARNING
OPEN CIRCUIT BEFORE REMOVING
COVER. INSTALL IN ACCORDANCE WITH
THIS USER’S MANUAL
Take care not to generate mechanical
sparking when access to the instrument and
peripheral devices in a hazardous location.
(5) Maintenance and Repair
WARNING
The instrument modication or parts replacement
by other than an authorized Representative of
Yokogawa Electric Corporation is prohibited and
will void the certication.
(6) Name Plate
● Name plate
● Tag plate for flameproof type
● Tag plate for intrinsically safe type
F0209r.ai
D
D
AFTER DE-ENERGIZING, DELAY 5 MINUTES
BEFORE OPENING.
WHEN THE AMBIENT TEMP. 65°C,
USE THE HEAT-RESISTING CABLES 90°C
No. KEMA 07ATEX0109
Ex d IIC T6...T4, Ex tD A21, IP6X
Enlcosure : IP66, IP67
TEMP. CLASS T6 T5 T4
MAX PROCESS TEMP.(Tp.) 85 100 120 °C
Tamb. -50(-15) to 75 80 75 °C
T85°C(Tamb.:40°C, Tp.:80°C),T100°C(Tamb.:60°C, Tp.:100°C),
T120°C(Tamb.:80°C, Tp.:120°C) Min.Tamb.:-40(-15)°C(for Dust)
No. KEMA 03ATEX1544 X
EEx ia IIC T4
IP66 and IP67
Tamb. -50(-15) to 60°C MIN. Tamb.:-40(-15)°C(for DUST)
MAX. PROCESS TEMP.(Tp.) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Ui=30V, Ii=200mA , Pi=0.9W, Ci=10nF, Li=0