  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 173
10-13
IM DR232-01E
10
Basic Settings (SET UP)
10.5 Setting Burn-out/Reference Junction Compensation
Setting Burn-out (BURN OUT)
Set the dot-printing position when burn-out occurs for each channel.
OFF: the burn-out function is not actuated.
DOWN: a dot is printed at the end of chart width on the minimum value setting side of the
recording span.
UP: Dot is printed at the end of chart width on the maximum value setting side of the recording
span.
The default setting is OFF.
Setting Reference Junction Compensation (RJC)
Set either the internal or the external compensation for each channel.
INT: internal compensation is performed.
EXT: external compensation is performed. For external compensation, set the voltage to be
compensated.
Voltage to be externally compensated: set in the range of -20000 to 20000 mV.
The default setting is INT.
Note
In order to make the set or selected contents effective, the contents must be registered. For details, see
page 10-23.
To return from the SET UP menu to the operation display mode, select END in the main menu. For
details, see page 10-23.