Power Supply
(Zone 2 only)
Transmitter
Supply
Nonhazardous Location
[Installation Diagram]
Hazardous Location
+ 
+ 
F0209.EPS
Ratings of the Power Supply as follows;
Maximum Voltage: 30 V
CENELEC ATEX Type of Protection “Dust”
[When combined with II 2G]
Follow the installation instructions described in 'b.
CENELEC ATEX Flameproof Type.'
• Applicable Standard: EN61241-0:2006, EN61241-
1:2004
• Type of Protection and Marking Code: II 2D Ex tD
A21 IP6X
• Maximum Surface Temperature:
T85°C (Tamb.: –40* to 75°C, Tp:85°C), T100°C
(Tamb.: –40* to 80°C, Tp:100°C), and T120°C
(Tamb.: –40* to 75°C, Tp:120°C)
*–15°C when /HE is specified.
Note 1. Installation instructions
The cable entry devices and blanking elements
shall be of a certificated type providing a level
of ingress protection of at least IP6X, suitable
for the conditions of use and correctly installed.
[When combined with II 1G]
Follow the installation instructions described in 'a.
CENELEC ATEX Intrinsically Safe Type.'
• Applicable Standard: EN50281-1-1:1997
• Type of Protection and Marking Code: II 1D
• Maximum Surface Temperature:
T65°C (Tamb.: 40°C), T85°C (Tamb.: 60°C), and
T105°C (Tamb.: 80°C)
Note 1. Installation instructions
The cable entry devices and blanking elements
shall be of a certificated type providing a level
of ingress protection of at least IP6X, suitable
for the conditions of use and correctly installed.
WARNING
OPEN CIRCUIT BEFORE REMOVING
COVER. INSTALL IN ACCORDANCE WITH
THIS USER’S MANUAL
Take care not to generate mechanical sparking
when access to the instrument and peripheral
devices in hazardous locations.
( III )
Tag plate for flameproof type
Tag plate for flameproof, intrinsically safe type,
type n protection, and Dust
F0298.EPS
D
T65C (Tamb.: 40C), T85C (Tamb.: 60C),and T105C (Tamb.: 80C)
AFTER DE-ENERGIZING, DELAY 10 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP. 70C, USE THE HEAT-RESISTING
CABLES90C. SEE USER'S MANUAL BEFORE USE.
SEE USER'S
MANUAL
Tamb. 40(15) TO 60C
TEMP.CLASS/MAX SURFACE TEMP.
T6/T85C
T5/T100C
T4/T120C
CAUTION
D
, Ex tD A21 IP6X T85C,T100C,T120C
ENCLOSURE:IP67
Tamb. 40(15) TO
KU21
Tamb. 40(15) TO 75C(T6 and T4)
AFTER DE-ENERGIZING, DELAY 10 MINUTES BEFORE OPENING.
WHEN THE AMBIENT TEMP. 70C, USE THE HEAT-RESISTING
CABLES90C. SEE USER'S MANUAL BEFORE USE.
40(15) TO 80C(T5)
CAUTION
KF21