Yokogawa EJA115 Manual de usuario Pagina 17

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 31
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 16
IM 01C22T01-01E
2-10
2. HART COMMUNICATOR OPERATION
(7) Bi-directional Flow Measurement
(a) Bi-dir mode enables selection of 50% output at an
input of 0 mmH
2
O.
F0218.EPS
(ENTER)
EJA:
Bi-dir mode
OFF
off
on
HELP SEND ESC ENTER
Call up the Bi-dir mode Display
Select on, and press ENTER (F4).
Press SEND (F2) to send the data
to the transmitter, then check to
confirm that disappears.
(SEND)
1. Device setup
4. Detailed setup
2. Signal condition
8. Bi-dir mode
SEND
Note: The measurement range changes to –3000 to 0 to
3000mmH
2
O (output 0% to 50% to 100%).
Note that LRV and URV are not changed.
Example: If measurement range is 0 to 3000mmH
2
O
(LRV = 0 mmH
2
O, URV = 3000 mmH
2
O)
(b) Combining Bi-dir mode with Xfer fnctn provides
a square root output computed independently for
0% to 50% output and for 50% to 100% output.
20 mA (100% display)
4 mA (–100% display)
Output mode “LINEAR”
LRV HRV
F0219.EPS
20 mA (100% display)
Low Cut
4 mA (–100% display)
Output mode “SQUARE ROOT”
LRV HRV
(8) Change Output Limits
The range of normal ouput is preset at factory from
5.0 to 110.0% unless otherwise specified or condi-
tioned, and the output is limited with these upper and
lower values. This output range can be changed, for
example, to meet the requirements of NAMUR, within
the settable range. Set the lower limit with AO lower
limit % and upper limit with AO Upper Limit %.
Settable range : 5.0 to 110.0 (%),
Upper limit > Lower limit
(9) Integral Indicator Display Mode
F0220.EPS
(ENTER)
EJA:
Display fnctn
Linear
Linear
Square Root
HELP SEND ESC ENTER
Select Square Root and press
ENTER (F4).
Press SEND (F2) to send data.
(SEND)
1. Device setup
4. Detailed setup
4. Display condition
2. Display fnctn
Example: Change from Linear to Sq root
Vista de pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 30 31