9.LIST OF PARAMETERS
No. Item Display Remarks No. Item Display Remarks
01 Model MODEL
02 Tag No. TAG NO
03 Self-check result SELF CHK
Display items
A Display1 DISPLAY1 B Display2 DISPLAY2
A01 Input value INPUT1 B01 Input value INPUT1
A33 Temporary memory 1 T1 B31 Integrating counter 1 COUNTER1
A34 Temporary memory 2 T2 B32 Integrating counter 2 COUNTER2
A54 Temporary memory 3 T3 B33 Integrating counter 3 COUNTER3
A55 Temporary memory 4 T4 B34 Integrating counter 4 COUNTER4
A54 Status STATUS *1 B60 Self-check result SELF CHK
A56 Rev. no REV NO
A58 MENU REV MENU REV
A60 Self-check SELF CHK
*1 The Status is displayed for service personnel to see history records.
Setting items
D Setting (I/O) SET(I/O) F Setting (communication) SET (COM)
D01 Tag no. 1 TAG NO.1 F01 Communication protocol PROTOCOL
D02 Tag no. 2 TAG NO.2 F02 Address ADDRESS
D03 Comment 1 COMMENT1 F03 Baud rate BAUD RATE
D04 Comment 2 COMMENT2 F04 Parity PARITY
D10 Range unit UNIT F05 Data length DATA LEN
D41 Pulse rate range PULSE RATE F06 Stop bit STOP BIT
D42 Pulse width type PULSE TYPE F07 Decimal point position of input
INPUT DEC PT
D43 Pulse width time PULSE TIME F60 Self-check result SELF CHK
D50 Input filter INPUT FILTER
D60 Self-check result SELF CHK
Adjusting items There are items not displayed depending on what output-2 is specified.
P Adjustment ADJUST1
P60 Self-check result SELF CHECK
10.MAINTENANCE
The product starts running immediately when the power is turned on;
however, it needs 10 to 15 minutes of warm-up before it meets the
specified performance.
For cleaning the instrument, use a soft and dry cloth.
10.1 Calibration Apparatus
Pulse generator (Yokogawa FG100 or the equivalent): 1
A counter(Yokogawa TC100 or the equivalent) or oscilloscope
(Yokogawa DL1540 or the equivalent): 1
A precision resistor (1k
⍀
, 1.6 k
⍀
): 1 each
6V battery: 1
10.2 Calibration Procedure
Connect the instruments as shown in Fig.10.1. First adjust the out-
put-1 signal and then the output-2 signal.
Produce a rectangular pulse of any frequency from the pulse genera-
tor to measure the value using a counter or oscilloscope. (Connect the
counter or oscilloscope as the broken line shown in figure 10.1.)
Next, connect the counter to the terminals 7 and 9, or terminals 2 and
5, then check that the frequency (input frequency x set rate) is output.
When using a oscilloscope, the wave shaping of output pulse can be
confirmed.
1011
3 2
1
4
56
789
1110 8
L+
N-
GND
9
7
Calibration for output-1
Calibration for output-2
4
3
1.6kΩ
1kΩ
6V
Battery
6V
Battery
+
-
+
-
5
2
4
3
1.6kΩ
1kΩ
+
-
+
-
Power supply
Pulse
Generator
Counter or
Osilloscope
Pulse
Generator
Counter or
Osilloscope
Fig. 10.1