IM 1E10C1-E
7-24
7. OPERATION VIA HART COMMUNICATOR
(9) Write Protection
Write protection function is used to prevent unauthorized data being written into the
device, ADMAG SE. This function becomes active by entering a password in “New
password”. Write protection status is released for 10 minutes by entering the password
in “Enable wrt 10min”.
When the write protection function is active, data setting changes in all parameters of
the ADMAG SE are inhibited and cannot be changed using either the HART Commu-
nicator or the ADMAG SE front panel key switches.
Setting Password
ABORT ENTER
3. New password
ADMAG SE :
Hot key
1 Span 1 m/s
2 Wrt protect menu
HELP SAVE HOME ENTER
1
2
ADMAG SE :
Enter new password to
change state of write
protect
HELP DEL ABORT ENTER
3
Call up “Wrt protect menu” in Hot key menu.
HELP SAVE
4
YZ /
3
Hot key 2. Wrt protect menu2. Wrt protect menu
Example: Set the password to “1 2 3 4
Select “Wrt protect menu”.
Select “New password”.
Set “1 2 3 4” and press ENTER (F4).
Set “1 2 3 4” again and press ENTER (F4).
Write protect” status changes from “No” to “Yes”.
Press OK (F4).
ABORT OK
ADMAG SE :
Change to New password
DEL DEL
5
F4
6
(OK)
ADMAG SE :
Wrt protect menu
1 Write protect No
2 Enable wrt 10min
3 New password
4 Software seal Keep
ADMAG SE :
Re-enter new password
within 30 seconds:
1234
1234
HELP DEL ABORT ENTER
ABORT
ADMAG SE :
It changed the state
of protection related
password
HELP DEL OK
F4
STU
1
VWX
2
YZ /
3
JKL
4
(ENTER)
F4
STU
1
VWX
2
YZ /
3
JKL
4
(ENTER)