11. PARAMETER LIST
Indicator Name R/W DescriptionData Range, Units
BRAIN
Terminal
Decimal
Point
Default
Value
— G00 4 - 20 SET R
G01 G01 4 - 20 LOW CUT 0 to 10% 0 W
G02 G02 4 - 20 LOW LMT –20 to 100% –20 W
G03 G03 4 - 20 H LMT 0 to 120% 120 W
— G60 SELF CHECK GOOD R
to
EEPROM DEFAULT
— H00 TEST R
H01 H01 TEST MODE (00) NORMAL
NORMAL
W
(01) TEST
H02 H02 OUTPUT VALUE –108 to 108 0 W
H03 H03 STATUS OUT (00) NORMAL
NORMAL
W
(01) CLOSED (ON)
(02) OPEN (OFF)
— H60 SELF CHECK GOOD R
to
EEPROM DEFAULT
— L00 OTHER R
L01 L01 TUNING (00) INHIBIT
ENABLE
W
(01) ENABLE
L02 L02 KEY 00 0 W
55 << TO “N” >>
— L60 SELF CHECK GOOD R
to
EEPROM DEFAULT
— N00 APPL SET
n01 N01 TOTAL / PULSE (00) NO DAMP DAMP W
(01) DAMP
n02 N02 OUTPUT MODE (00) ON ON W
(01) OFF
n03 N03 RATE LIMIT 0 to 10% 5 W
n04 N04 DEAD TIME 0 to 15 s 0 W
n05 N05 POWER SYNCH (00) YES YES W
(01) NO
n07 N07 EMPTY PIPE (00) ALARM NO W
(01) NO ALARM ALARM
— N60 SELF CHECK GOOD
to
EEPROM DEFAULT
Current output functions.
Sets low cut width of current output.
Sets low limit of current output.
Sets high limit of current output.
Displays self-check result.
Output circuit test mode.
Selects test mode.
Sets test output value.
Selects test status output.
Displays self-check result.
Data setting restriction functions.
Inhibits or enables data setting with
indicator switches or BT200.
Enables “N” items display. Sets “55”
for access to “N” items.
Displays self-check result.
Special application.
Selects whether instantaneous flow rate
values or damped ones are used for
totalization and pulse output.
Selects active level of status output set
in “B10.” (except “ALARM OUT”)
Sets the level to reduce output fluctua-
tion.
Sets dead time to reduce output fluctua-
tion. When “0” is set, “Rate limit” is not
available.
Selects synchronization between excita-
tion frequency and power frequency.
Selects alarm effectiveness when flow
tube is not filled with fluid.
Displays self-check result.