Yokogawa DO402 Dissolved Oxygen Analyzer Manual de usuario Pagina 92

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 98
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 91
IM 12J05D02-01E
11-4 Appendix 2
4/4
QIS 12J05D02-01E
3.7 Atmospheric Pressure Indication Check
Following Section 3.6, press the [ENT] key until the message display shows “PRESS.”
Check that the data display indication at this time is within the range from 900 (hPa) to
1100 (hPa).
Press the [ENT] key. The message display will show “READY.”
Continue to press the [ENT] key to return to normal measurement mode.
CONTACT
FAIL S3 S2 S1
3 2 1 32 71 72 51 52 41 42 31
11
T
1
12
T
2
61
62 65 66
15
RE
13
IE
17
+
-
+
-
mA 1 mA 2
Figure 1 Testing Circuit and Test Equipment
DO402G Converter
Decade
Resistance Box

Power Supply
Standard Current
Generato
r
+
-
SW1
+
-
DC Ammeter
Current Output 1
300
1%
+
-
DC Ammeter
300
1%
Current Output 2
Vista de pagina 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 98

Comentarios a estos manuales

Sin comentarios