IM 12D08B02-01E
Appendix 3-7
3/3
QIS 12D08B02-21E
Table 3
Message display Decade box 2 resistance Data display
RES.1
10 : 10.00 ±0.05 :
RES.2
100 : 100.0 ±0.5 :
RES.3
1 k: 1.000 ±0.005 k:
RES.4
10 k: 10.00 ±0.05 k:
RES.5
100 k: 100.0 ±0.5 k:
Figure 1 Testing Circuit and Test Equipment
SC202S
SUPPLY SENSOR
+
-
G
11 12 1413 15 16
24V DC
+
100 :
DC source
+
-
-
DC
Decade
Box 1
Note 1: Cable connected to sensor input should be conductivity detector cable of
length 2.1 ±0.1 m. Connect pins 13 and 14, also 15 and 16, to Decade box 2
terminals. Use shielded cable, and connect shield to pin 14.
Note 1
Decade
Box 2
Ground
Milli-
ammeter