23
IM 12D06D05-01E
5
MENU STRUCTURECOMMISSIONING
Menu Parameter Default Range
values min. max.
Calibration Airadjust 100.00µS 0µS 100µS
c.c.high 120% 100% 120%
c.c.low 80% 80% 100%
Stabilizationtime 5s 2s 30s
Calib. interval 250 days 1 day 250 days
Concentr.Table Table Seeappendix
Enter
Measurement setup
Temperature settings
Temp. Compensation
Calibration settings
Concentration
Measure Conductivity only
Configure sensor
Concentration measurement isonly possible if“measure” inthe
“Configuresensor” menu is setto “conductivity+concentration”
or“Concentrationonly”.