B. Calculation of temperature coefficient factor
(with two known conductivity values at different temperatures)
Measure the conductivity of the liquid at two temperatures, one below the reference and above
thereferencetemperaturewiththetemperaturecoefficientsetto0,00%/°Candusethefollowing
equationtocalculateatemperaturecoefficient(α).
K
T
K
ref
=
1+α ( T - T
ref
)
K
2
- K
1
α
=
K
1
( T
2
- T
ref
) - K
2
( T
1
- T
ref
)
K
1
K
2
K
ref
=
=
1+α ( T
1
- T
ref
)
1+α ( T
2
- T
ref
)
K
1
(1 + α ( T
2
- T
ref
)) = K
2
(1 + α ( T
1
- T
ref
))
K
1
+ α ( T
2
- T
ref
) - K
2
+ α ( T
1
- T
ref
) = K
2
- K
1
Where T
1
, T
2
:liquidtemperature(°C)
K
1
: conductivity at T
1
(°C)
K
2
: conductivity at T
2
(°C)
Figure 11-1. Conductivity
Calculation example
Calculate the temperature co-efficient of a liquid from the following data.
Conductivity124.5µS/cmataliquidtemperatureof18.0°Canda
conductivity147.6µS/cmataliquidtemperatureof31.0°C.
Substitutingthedataintheaboveformulagivesthefollowingresult.
147.6 - 124.5
x 100= 1.298 %/C
α =
124.5(31.0 - 25) - 147.6(18.0 - 25)
SetthetemperaturecoefficientintheISC450Gconverter.