Yokogawa PC-Based MX100 Manual de usuario Pagina 114

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 133
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 113
4-26
IM MX100-01E
4.9 10-CH, High-Speed Digital Input Module
(MX115) Specifications
Style number: S1(-D05), S2(-D24)
Input type: -D05: contact (non-voltage contact, open collector), and LEVEL
(5-V logic)
-DI24: LEVEL (24 V logic)
Number of inputs: 10
Input format: -D05: Pull-up at approximately 5 V/approximately 5 kΩ, no
insulation between channels
-D24: No isolation between channels:
Measurement interval: Select 10 ms, 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s,
20 s, 30 s, or 60 s
Filter: Measurement interval of 5 s or less: Uses the widest ON/OFF
width during the detection period (approximately 75% to 90% of
the measurement interval).
Measurement interval of 5 s or more: Uses the widest ON/OFF
width during approximately 4.5 s
Minimum detection pulse width:
Twice the sampling interval or more
Input threshold level:
-D05:
Contact (non-voltage contact, open collector):
ON at 100 Ω or less and OFF at 100 kΩ or more
LEVEL (5-V logic): OFF at 1 V or less and ON at 3 V or more
-D24:
LEVEL (24 V logic): OFF at 6 V or less and ON at 16 V or more
Hysteresis width: -D05: Approximately 0.1 V
-D24:
Approximately 1.5 V
Contact/transistor rating:
Contact with a rating of 15 VDC or greater and 30 mA or greater
Vce and Ic are transistors with ratings of 15 VDC or more, and 30
mA or more, respectively.
Maximum input voltage: -D05: ±10 V
-D24: Approx. ±50 V
Insulation resistance: 20 MΩ or more (500 VDC) between the input and ground
Withstand voltage: 2300 VAC (50/60 Hz) for one minute between input and earth
Maximum common-mode voltage: 250 V
AC rms (50/60 Hz)
Terminal type: Clamp
Power consumption: Approximately 1.5 W
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal cover)
Weight: Approximately 0.5 kg
External Dimensions
Unit: mm
57
131
150.8
If not specified, the tolerance is ±3%. However, in cases of less than 10 mm, the tolerance is ±0.3
mm.
Vista de pagina 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 132 133

Comentarios a estos manuales