Index
1.5 Functions of the 30-CH, Medium-Speed DCV/
TC/DI Input Module
This module is equipped with 30 inputs and measures DC voltage, Thermocouple, and
digital input (DI) at a minimum measurement interval of 500 ms.
It takes up three modules worth of space when attaching to the base plate.
Terminal
cover
Input terminal
(clamp terminal)
Terminal
cover
Input terminal
(M3 screw terminal)
MX110-VTD-L30/H3MX110-VTD-L30
Input Type and Measurement Range
The following measurements are possible.
Input Measurement Range Type Rated Measurement Range
DC voltage
20 mV –20.000 to 20.000 mV
60 mV –60.00 to 60.00 mV
200 mV –200.00 to 200.00 mV
2 V –2.0000 to 2.0000 V
6 V –6.000 to 6.000 V
20 V –20.000 to 20.000 V
100 V –100.00 to 100.00 V
Thermocouple
R
0.0 to 1760.0°C
S
B 0.0 to 1820.0°C
K –200.0 to 1370.0°C
E –200.0 to 800.0°C
J –200.0 to 1100.0°C
T –200.0 to 400.0°C
N 0.0 to 1300.0°C
W 0.0 to 2315.0°C
L –200.0 to 900.0°C
U –200.0 to 400.0°C
KPvsAu7Fe 0.0 to 300.0K
DI
Level Vth = 2.4 V
Contact input
ON: 1 kΩ or less,
OFF: 100 kΩ or more
(parallel capacitance: 0.01 µF
or less)