Explanation of the Advanced Security Function
IM 04L41B01-05EN
1-7
Recording Conditions of Event Data
Item Description
Channel type Same as display data.
Sampling interval Choices are available in the range of 25 ms to 30 min. You cannot choose a
sampling interval that is higher than the scan interval.
File generation A file is generated when the set data length is reached.
Files are also generated in these cases:
• When you generate a file manually.
• When sampling stops (memory stop).
• When a file is generated through the use of the event-action feature.
• After the DX recovers from a power failure
Mode Free (always recording)
Pressing the START key starts recording. Pressing the STOP key ends
recording.
Creating Files through Key Operation
You can use key operations to generate les.
For operating instructions, see section 6.4 in the User’s Manual, “Starting/Stopping the
Recording and Saving the Measured data.”
1.2 Recording and Saving Data
Time
File File File Adding data
Time
Previous files
key operation
FileFile File
Current file