Calibration
8.1 Operating Procedures on the Calibration
Screen
Displaying the calibration screen
Calibration is done using the Software Configurator.
After selecting the project, click the Software Configurator from the launcher software or
right-click the launcher icon on the task bar and select Start Software Configurator.
The Software Configurator starts.
Clicking the Calibration tab displays the calibration screen, below.
Calibration tab
Connect button
Calibration procedures
There are two ways to perform calibration.
Auto: Apply the specied voltage to the terminal displayed on the screen, and
calibrate all channels based on that input signal.
Manual: Calibrate by setting a calibration value for each channel.
Chapter 8 Calibration