Yokogawa DA100 Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 105
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
4-13
IM DA100-11E
Overview and Specifications of Ethernet Interface
4
4.9 Setting the Timeout
A PC connected to the DA100 for a certain period of time is automatically disconnected.
ET Sets the timeout.
Mode Operation mode
Syntax ETp1<terminator>
p1 Timeout time
0 No timeouts (Default setting).
1 Disconnect after one minute without any access
2 Disconnect after 15 minutes without any access
3 Disconnect after 30 minutes without any access
4 Disconnect after one hour without any access
5 Disconnect after two hours without any access
6 Disconnect after ten hours without any access
Description This command is common to both ports, 34150 and 34151.
Note
The timeout set with the ET command is used for disconnecting a PC that stays connected for a certain period of time
without any operation.
KeepAlive is used for disconnecting a connection when the physical connection with the PC is disconnected or the PC goes
down.
ET command does not support sub-delimiters.