3.13 Managing the Input Calibration Interval (/CC1
option; release numbers 4 and later)
Manage the input calibration interval to make sure that calibration is performed regularly.
For a description of the function, see section 1.1.
Setup Screen
Press MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic
setting mode), and select the Menu tab > Set Calibration.
Setup Items
Set Calibration > Use/Not
Select whether to Use or Not use calibration management.
When you select Use, the Set Calibration tab appears in setting mode.
Set Calibration > Notification (DayBfr)
You can specify how many days before the calibration due date you want to display
the calibration notification screen. You can set the number of days to a value between
1 and 10.
Set Calibration > Renotification
You can specify the period at which to display the calibration notification screen. The
calibration notification screen will continue to appear until calibration is completed.
Setting Description
10min The screen will appear every 10 minutes.
30min The screen will appear every 30 minutes.
1h The screen will appear every hour.
8h The screen will appear every 8 hours.
24h The screen will appear every 24 hours.