  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 111
4-12
IM MV1000-01E
Explanation
Operations That Begin at the Same Time as Memory Sampling (Memory
start)
Waveform display updating on the trend display
Reporting (/M1 and /PM1 options)
Computation functions (/M1 and /PM1 options), if they are congured to do so.
For details, see section 10.4.
Operations That Stop at the Same Time as Memory Sampling (Memory
stop)
Waveform display updating on the trend display
Reporting (/M1 and /PM1 options)
Computation operations (/M1 and /PM1 options), if they are being performed.
Save Problems
The following problems may arise when the MV is continually accessing internal or
external memory. If these problems arise, the storage media access indicator will light
frequently.
External storage media le save error
Access timeout
These problems can be dealt with in the following ways.
If you are using the event action function and saving small consecutive les,
increase the size of the les being saved.
If you are saving multiple les to the same directory on an external storage
medium, change the name of the Save directory, making sure that the number of
les in the directory does not exceed 1000.
If data recording and display are using up resources (for example if you are
recording on multiple channels using a fast sampling rate, and displaying four trend
displays on the 4-panel display), use a slower sampling rate or change the display
.
4.4 Starting/Stopping Recording and Saving Measured Data