2-24
SM MX100-01E
2.6.6 Between-Channel Error Test
Procedure
1.
Connect the I, A, B, and C terminals on all channels to a voltage generator or
variable resistor, then confirm that the measured values on each channel meet
the criteria.
Set the measurement interval on the module under test to 500 msec.
2.
Confirm that the between-channel error vales are less than the criteria given
below.
Range Input Criteria
20 mV Short input H to L 0.000 mV ±0.005 mV
20 Ω 20 Ω 20.000 Ω ±0.017 Ω
2.6.7 DI Contact Input Test
Connect the resistances below between the A and B terminals on any two channels, then
confirm that the criteria are met.
Set the measurement interval on the module under test to 500 msec.
Range Load Criteria
DI CONTACT 2 kΩ 1
DI CONTACT Connect 200 kΩ + 0.022 μF in parallel 0 or 20 kΩ