Items to Be Checked
Confirm that all channels under test satisfy the specifications in the table below.
Also confirm that the output from channels not under test meet the criteria for 0 V in the
table below.
<MX100>
Voltage Output Criteria
Setting Standard Value Specification
-10 V (20000) -10 V ±20 mV
0 V (0) 0 V
+5 V (10000) +5 V
+10 V (20000) +10 V
Current Output Criteria
Setting Standard Value Specification
+1 mA (500) 1 mA ±40A
+10 mA (5000) 10 mA
+20 mA (10000) 20 mA
+Over (20000) 22 mA or more 22 mA or more
The parentheses ( ) indicate the count value when entering settings with the MX100/
DARWIN API.
<MW100>
Voltage output criteria
Setting Standard Value Specification
-10 V (-10000) -10 V ±20 mV
0 V 0 V
+5 V (5000) +5 V
+10 V (10000) +10 V
Current Output Criteria
Setting Standard Value Specification
+1 mA (1000) 1 mA ±40A
+10 mA (10000) 10 mA
+20 mA (20000) 20 mA
+Over (25000) 22 mA or more 22 mA or more
2.9 Tests of the 8-CH Analog Output Module