2-51
SM MX100-01E
Testing
1
2
3
4
5
6
7
<MW100 Settings>
Item Setting Remarks
Output Type PWM
Output method COM Command output
Output range 1 ms
Span Need not specify
Pulse interval 10
Preset value 0.000%
Reference channel Not required
Output Operation Setting
Item Setting
Power ON PRESET
Error occurence PRESET
Criteria (for the MX100 and MW100)
Change the settings following the steps in the table below, then confirm that the results
satisfy the criteria for voltage.
Step External Power Supply PWM Output Setting Voltage Criterion
1 5 V 100.000 % 3.8 V or more
2 12 V 100.000 % 0.04 V or less
3 5 V 100.000 % 0.04 V or less
4 OFF 100.000 %
5 5 V 100.000 % 3.8 V or more
2.10 8-CH, Medium-Speed PWM Output Module