IM MW100-01E
5.13 Operations Common to the 8-CH Medium-
Speed Analog Output Module and the 8-CH
Medium Speed PWM Output Module (MX120)
Settings Related Specifications (by Module)
Setting channel
(Module)
Setting Contents Settings Remarks
Output Channel
(AO, PWM)
Span Setting
range
AO(V) –10.000 to 10.000V
The output range
for arbitrary output
is within the range
on the left
AO(mA) 0.000 to -20.000 mA
PWM 0.000 to 100.000%
Preset value
Setting
range
AO(V) –11.000 to 11.000V
AO(mA) 0.000 to 22.000 mA
PWM 0.000 to 100.000%
Setting span (minimum and maximum)
specified in reverse
Yes
Setting span (minimum, maximum)
same value specification
No
Overview of Output Operation When Setting Holding of Previous Value of
Transmission Output
Measurement
start
Measurement
stop
(or Transmission output
execute disabled)
Measurement
restarts
(or Transmission output
execute enabled)
Previous value
(held)
Previous value
Power ON
Output value
Value held