Yokogawa Standalone MW100 Manual de usuario Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 305
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
1-41
IM MW100-01E
Explanation of Functions
1
2
3
4
5
App
Index
1.6 Functions of the 30-CH, Medium-Speed DCV/
TC/DI Input Module
This module allows up to thirty inputs of DC voltage, thermocouple, and digital input (DI)
at a shortest measurement interval of 500 ms. It takes up three modules worth of space
when attaching to the base plate.
Input terminal
(M3 screw terminal)
Terminal
cover
Input terminal
(clamp terminal)
Terminal
cover
MX110-VTD-L30/H3MX110-VTD-L30
Measurement Input Types
Measurement Input Type Notation
No measurement SKIP
DC voltage VOLT
Thermocouple TC
DI DI
Remote RJC RRJC
Measurement Range
DC Voltage
Measurement Range Notation Rated Measurement Range
20 mV 20 mV –20.000 to 20.000 mV
60 mV 60 mV –60.00 to 60.00 mV
200 mV 200 mV –200.00 to 200.00 mV
2 V 2 V –2.0000 to 2.0000 V
6 V 6 V –6.000 to 6.000 V
20 V 20 V –20.000 to 20.000 V
100 V 100 V –100.00 to 100.00 V
60 mV (high resolution) 60 mVH 0.000 to 60.000 mV
6 V (high resolution) 6 VH 0.0000 to 6.0000 V
1 V 1 V –1.0000 to 1.0000 V
Thermocouple
Measurement Range Notation Rated Measurement Range
Type-R R 0.0 to 1760.0°C
Type-S S 0.0 to 1760.0°C
Type-B B 0.0 to 1820.0°C
Type-K K –200.0 to 1370.0°C
Type-E E –200.0 to 800.0°C
Type-J J –200.0 to 1100.0°C
Type-T T –200.0 to 400.0°C