Yokogawa 2-Wire Dual Channel Transmitter/Analyzer FLXA21 Manual de usuario Pagina 37

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 246
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 36
<2. WIRING AND INSTALLATION>
2-10
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
2.5.1 Wiring the pH/ORP sensor
n pH Measurement
Conventional pH sensors are connected to the module as follows:
11 Temperature
12 Temperature
13 Reference
14 Solution ground
15 Glass (measure)
16 Shield
17 Shield
FLXA21
REFTC PHLiquid
Earth
In addition to the wiring of the sensor, insure that a jumper for low-impedance sensor inputs is
installed. The jumpers can be found on the plastic sensor module cover and can be stored in the
lower level module wiring cover.
• pHGlassElectrodeisahighimpedancesensorinput
• StandardreferenceelectrodesandanORP/REDOXelectrodearelowimpedancesensor
inputs
• Specialelectrodesusing2glasssensor(example:Pfaudler,SC24V)donotneedjumpers.
Terminals15-16areidentiedasinput1(HighImpedance)andterminals13-17aredenedas
input 2 (Low Impedance). For conventional pH sensors, the jumper is placed as illustrated:
16
15191713181412
11
PH
Input 1
Input 2
Glass sensor on Input 1.
Reference sensor on Input 2.
Vista de pagina 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 245 246

Comentarios a estos manuales

Sin comentarios