<5. CALIBRATION OF pH/ORP>
5-5
IM 12A01A02-01E 5th Edition : Oct. 31, 2013-00
5.1.3 Sample calibration
A sample calibration is a single-point calibration for only the zero (asymmetric). It adjusts the
recordedreadingtoacollectedsamplevalue.Press[TakeSample]torecordacollectedsample
valueinmemory.Re-entertheSampleCal.screenandpress[Startcalibration]toperforma
sample calibration. This updates the recorded data.
NOTE
Whenasensororanelectrodeisexchangedorreplaced,sensorwellnessdatashouldbereset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to
thegure3.9.)
5.2 Temperature calibration
For the most accurate measurements, it is important to have a precise temperature
measurement. Measure the temperature with a high-precision thermometer and adjust the
sensor reading accordingly. For best accuracy, this should be done as near to the normal
operating temperature as possible.
5.3 ORP calibration (rH calibration)
ThecalibrationmodesforORPorrHare“Manual”and“Sample”.Noautomaticcalibration
feature is available in an ORP or rH calibration.
Calibration is performed stepwise. Follow the prompts displayed on the screen.
Astabilitycheckismadeateachmeasurementpoint.Proceedtothenextsteponlyafterthe
reading has stabilized.
NOTE
Whenasensororanelectrodeisexchangedorreplaced,sensorwellnessdatashouldbereset.
When a sensor is replaced, the replacement can be recorded manually into a logbook. (Refer to
thegure3.9.)