Yokogawa Value Series FX1000 Manual de usuario Pagina 10

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 206
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 9
1-2
IM 04L21B01-17EN
Modbus Server
A Modbus client device can carry out the following operations on the FX that is
operating as a Modbus server device.
Load data from measurement channels and computation channels
*
(using the input
register)
Load communication input data
1
(using the hold register)
Write communication input data
1
(using the hold register)
Start and stop recording, write messages, and perform other similar operations
(using the hold register)
Load the recording start/stop conditions and other conditions (using the hold
register)
1: /M1, /PM1, and /PWR1 options
For details on the Modbus function codes that the FX supports, see section 6.3.
For a description of the settings required to use this function, see section 1.9.
Modbus client device
FX
Modbus server
Ethernet
MW100
1.1 What You Can Do with the FX1000
Vista de pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 205 206

Comentarios a estos manuales

Sin comentarios