2-21
IM 04L21B01-64EN
Configuring the FX1000
2
Display
Logging
• Trend interval [/div]
This is the trend interval. Select the time corresponding to 1 division of the time axis
on the trend display from below: You cannot set a T-Y interval that corresponds to
a sampling interval that is faster than the scan interval. See the table under “Save
Interval” below.
• Save Interval (when recording display data)
Select the size of a record data le. The recorded data is divided by the le size
specied here. The available settings vary depending on the number of memory
sampling channels and the T-Y interval setting.
T-Y interval 15 s
1
30 s 1 min 2 min 5 min
Selectable range
of auto save
interval
500 ms 1 s 2 s 4 s 10 s
Selectable save
interval values
10 min to 3
days
10 min to 7
days
10 min to 14
days
10 min to 14
days
10 min to 31
days
T-Y interval 10 min 15 min 20 min 30 min 1 h
Selectable range
of auto save
interval
20 s 30 s 40 s 1 min 2 min
Selectable save
interval values
10 min to 31
days
10 min to 31
days
1 hour to 31
days
1 hour to 31
days
1 hour to 31
days
T-Y interval 2 h 4 h 10 h
Selectable range
of auto save
interval
4 min 8 min 20 min
Selectable save
interval values
2 hours to
31 days
4 hours to
31 days
8 hours to
31 days
1 Selectable on the FX1002 and FX1004