Configuring the FX1000
1
2
3
8
Index
Detail Setting
Tag
Settings Description
Tag Displays tags. Channel numbers are displayed for channels that do not have
tags assigned to them.
Channel Displays channel numbers.
Language
Select the display language
Decimal Point Type
You can set the decimal point type for the display and les saved in text format. You
can select [Point] or [Comma].
Batch
Select [ON] to use the batch function.
Digit of lot number
Select the number of digits of the lot number from 4, 6, or 8. Select [OFF] to disable
the lot number.
Auto increment
ON Automatically sets the lot number of the next measurement to “the lot number of
the current measurement + 1.”
Partial
Turn Partial [ON] (partially expand) or [OFF] (do not partially expand).
Trend Rate Switching
ON Enables the function that switches the trend interval while the memory sampling
is in progress. The “Second interval [/div]” item is displayed in the setting mode.
When [Trend Rate Switching] is set to [ON], you cannot set [Data Kind] under [Environment]
- [Basic Environment] in the [Basic setting] tab to [E+D].
Write Group
Settings Description
Common Write the message to all groups.
Separate Write the message to the displayed group.