4-8
IM WX103-01E
Example of the M4 screw terminal
Start calibration
When you click the Calibrate button, a message “Please wait and do not perform any
other operation.” appears.
5. Repeat steps 1 to 4 until all calibrations are completed.
6. From the F
ile menu, choose Write
.
The calibration values are written to the input module.
Note
During calibration, do not perform any other operations (particularly turning OFF the MX100 or
dropping the connection). Doing so may cause the MX100 to malfunction.
4.2 Calibration Procedure