<10. Parameter Summary>
10-10
IM 01C27C01-01EN
Table 10.2 Diagnostic Status Detail[0]
Bit Diagnostic Status Detail Description
Diagnostic Status
assignment bit
NAMUR
DiagnosticDetail_1
31 FC_SENSOR_FAIL C sensor frequency error Bit26 F
30 FR_SENSOR_FAIL R sensor frequency error Bit26 F
29 CAP_T_SENSOR_FAIL Capsule temperature sensor failure Bit26 F
28 CAP_EEPROM_FAIL Capsule EEPROM failure Bit26 F
27 CAP_EEP_IRREGULAR CAP EEPROM version not correct Bit26 F
26 AMP_T_SENSOR_FAIL Amplier temperature sensor failure Bit27 F
25 AMP_EEPROM_FAIL Amplier EEPROM failure Bit27 F
24 AMP_EEP_IRREGULAR AMP EEPROM version not correct Bit27 F
22 G_A_COMM_FAIL G/A failure Bit27 F
21 FC_UNOSC_FAIL C sensor oscillation stop failure Bit26 F
20 FC_DELTA_T_FAIL C-side deltaT circuit failure Bit27 F
19 FR_DELTA_T_FAIL R-side deltaT circuit failure Bit27 F
18 WL_AD_FAIL Battery voltage not detected (AMP failure) Bit27 F
17 FR_UNOSC_FAIL R sensor oscillation stop failure Bit26 F
15 DP_OUTSIDE_LIMIT Pressure outside of specied range Bit23 O
14 SP_OUTSIDE_LIMIT Static pressure outside of specied range Bit23 O
13 CAPT_OUTSIDE_LIMIT Capsule temperature outside of specied range Bit23 O
12 AMPT_OUTSIDE_LIMIT Amplier temperature outside of specied range Bit23 O
7 DP_OUTSIDE_RANGE Input pressure setting outside of range Bit22 O
6 SP_OUTSIDE_RANGE Static pressure setting outside of range Bit22 O
DiagnosticDetail_2
31 AI1_OUT_OF_SERVICE AI1 O/S mode Bit24 C
30 AI2_OUT_OF_SERVICE AI2 O/S mode Bit24 C
29 AI3_OUT_OF_SERVICE AI3 O/S mode Bit24 C
28 AI1_SIMULATION_ACTIVE AI1 simulation mode Bit17 C
27 AI2_SIMULATION_ACTIVE AI2 simulation mode Bit17 C
26 AI3_SIMULATION_ACTIVE AI3 simulation mode Bit17 C
20 DP_TRIM_SPAN_OUTSIDE Pressure span adjustment variable outside of range Bit25 C
19 DP_TRIM_ZERO_OUTSIDE Pressure zero adjustment variable outside of range Bit25 C
15 SP_TRIM_SPAN_OUTSIDE Static pressure span adjustment variable outside of
range
Bit25 C
14 SP_TRIM_ZERO_OUTSIDE Static pressure zero adjustment variable outside of
range
Bit25 C
13 WL_LOWBAT_ALM Low battery Bit19 M
12 CRITICAL_LOWBAT
*1
Detect a decrease in the lowest drive voltage. Bit20 M
10 WL_DEEPSLP_ALM Deep sleep due to low battery Bit20 M
8 LCD_OUTSIDE_LIMIT LCD display outside of specied range Bit25 C
*1: Applicable for amplier housing code 8 or 9.
*2: Applicable for amplier housing code 7.
F1001.ai
AI algoritm
MODE Target
from operator
from operator
from operator
from transducer
SIMULATE
On
Off
AI
algorithm
Linearization
scaling,
filter
Alert
report
HiHi etc
PV
Auto
O/S
Man
Publish
Value
& Status
PV
PV
Value
& Status
Value
& Status
Value
& Status
Mode and
PV.Status handling
SWITCH
SIMULATE
_VALUE
Figure 10.1 Example schema of analog input object