<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>
9-1
IM 01E30D01-01EN
9.
OPERATION VIA HART CONFIGURATION TOOL (HART 7)
ThisChapterdescribesthedevice(AXR)operation
proceduresusingaHARTcongurationTool(such
as FieldMate).
For more details regarding the operations for
theHARTcongurationtool,refertotheHART
congurationtooloperationsmanual.
Note: HART is a registered trademark of the HART
Communication Foundation.
9.1 HART Protocol Revision
For the models with the output signal code “-J”,
HART protocol revision 5 or 7 is selectable. The
protocolrevisionissetasspeciedintheorder.
Howtoconrmprotocolrevisionisshownbelow.
Therearetwowaystoconrmtheprotocolrevision
set to the transmitter.
(1)Conrmationbythenameplate
The HART protocol revision at the time of
shipment is shown by the last number at the
serial number column of the name plate.
42
F0901.ai
XXX…XX
5: HART 5
7: HART 7
(2) ConrmationbyusingHARTcongurationtool
Follow the procedure below.
1)Connectthecongurationtooltothedevice.
2) Call up “Review4” display.
[Root Menu]→Review→Review4
3) HART protocol revision is displayed on the
“Universal rev” column.
Protocol revision supported by HART
congurationtoolmustbethesameorhigher
than that of the device.
Protcol revision
supported by HART
c
ongurationtool
5 7
Protocol revision of
the device
5
7 ×
:
Communication OK
×: Communication NG
• In this User’s Manual, HART protocol
revision 5 and 7 are described as HART 5
and HART 7 respectively.
• HART 7 communication is supported by
FieldMate R2.04 or later.
When the output signal code of the device is “-J”,
HART protocol revision can be changed. Refer
to subsection 9.8.5 about the procedure of the
revision change of HART 5 and HART 7.
9.2 Device Description (DD) on
aHARTCongurationTool
and AXR Device Revision
9.2.1 DD Revision and Device Revision
BeforeusingtheHARTCongurationTool(such
asFieldMate),conrmthattheDD(Device
Description)oftheAXRisinstalledinthe
CongurationToolbeforeusing.
• AXR:HART7
• Device type: 0x3757
• Device Revision: 10
• DD Revision: 1
TheDDrevisionsforAXRandCongurationTool’s
canconrminaccordancewiththefollowingsteps.
If the correct DD is not installed in the HART
Comentarios a estos manuales