<3. INSTALLATION>
3-7
IM 01E30D01-01EN
Table 3.3.3 Wafer Type Tightening Torque Values for Metal Piping and Permeable Fluids
TighteningtorquevaluesforPFAliningtype(N-m/{kgf-cm}/[in-lbf])
Gasket types
w
ithinowtube
No gasket (standard)
Gasket types
for user’s
ange
PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
Flange ratings
Size
mm (inch)
JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
25 (1.0) 34.9 to 40.1 / {355.9 to 408.9} / [308.9 to 354.9] 35.2 to 40.1 / {358.9 to 408.9} / [311.5 to 354.9] 32.3 to 37.1 / {329.4 to 378.3} / [285.9 to 328.4]
40 (1.5) 53.5 to 61.5 / {545.5 to 627.1} / [473.5 to 544.3] 54.2 to 61.5 / {552.7 to 627.1} / [479.7 to 544.3] 56.4 to 61.5 / {575.1 to 627.1} / [499.2 to 544.3]
3.3.2 Nominal Diameter 50 mm (2.0 in.) to
200 mm (8.0 in.), Wafer Type
Useboltsandnutsincompliancewiththeange
ratings. When stud-type through-bolts are used,
be sure the outside diameter of the shank is
smaller than that of the thread ridge. Be sure to
choose a gasket with inner and outer diameters
that does not protrude inside the piping (refer to
Table 3.3.9). If the inner diameter of the gasket is
too large, or outer diameter of the gasket is too
small,uidleakagemayresult.
(1) Mounting Direction
Mounttheowmetersothattheowdirectionofthe
uidtobemeasuredisinlinewiththedirectionof
thearrowmarkontheowmeter.
If it is impossible to match the direction of
the arrow mark, the direction of the electrical
connection can be changed. Refer to Section
10.1 to do this properly.
Incasetheuidbeingmeasuredowsagainst
the arrow direction, refer to this parameter J20:
Flow Direction in this user’s manual.
(2) Mounting Centering Devices
Tomaintainconcentricityoftheowmeterwiththe
pipes, install centering devices. From the process
piping side, pass two through-bolts through the
four centering devices (two for each bolt) and
the adjacent two holes (the lower two holes for
h
orizontalmounting)ofbothoftheanges(refer
to Figure 3.3.2). Use the appropriate centering
devices according to the nominal diameter and the
angeratings.Thecenteringdevicesareengraved
with an identifying character. Use the appropriate
oneswhichmeettherequiredspecicationsby
referring to Tables 3.3.7.
(3) Positioning the Flowmeter
PositiontheowmetersothattheMini-angesand
the centering devices come in close contact with
each other. Be careful to prevent the four centering
devices from coming into contact with the housing.
If stud-type through-bolts are used, position them in
such a way that the four centering devices come in
contact with the bolt threads (refer to Figure 3.3.2).
Pass the other through-bolts through from the
process piping side.