Yokogawa ADMAG AXR Manual de usuario Pagina 182

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 194
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 181
<14. EXPLOSION PROTECTED TYPE INSTRUMENT>
14-5
IM 01E30D01-01EN
WARNING
IftheAXRismountedinanareawherethe
use of EPL Db equipment is required, it shall
be installed in such a way that the risk from
electrostatic discharges and propagating brush
dischargescausedbyrapidowofdustis
avoided.
(2) Electrical Connection
The type of electrical connection is stamped near
the electrical connection port according to the
following codes.
ISO M20x1.5 female M
ANSI 1/2NPT female N
F1401.ai
Screw Size Marking
(3) Installation
WARNING
• Groundingresistanceof100Ωorlessis
necessary.
When optional code A is selected, grounding
resistanceof10Ωorlessshallberequired.
All wiring shall comply with local installation
requirements and local electrical code.
In hazardous locations, the cable entry
devicesshallbeofacertiedIECEx
ameprooftype,suitablefortheconditionsof
use and correctly installed.
Unused apertures shall be closed with
suitableameproofcertiedblanking
elements.(Theplugattachediscertiedas
theameproofandIP66orIP67asapartof
this apparatus.)
In case of ANSI 1/2 NPT plug, ANSI
hexagonal wrench should be applied to
screw in.
The grounding terminals are located on the inside
and outside of the terminal area.
Connect the cable to grounding terminal in
accordance with wiring procedure 1) or 2).
F1404.ai
1) Internal grounding
terminal
2) External grounding
terminal
Figure 14.4.1 Wiring Procedure for Grounding
Terminals
(4) Operation
WARNING
After de-energizing, delay 5 minutes before
opening.
Take care not to generate mechanical
spark when access to the instrument and
peripheral devices in hazardous locations.
(5) Maintenance and Repair
WARNING
Electrostatic charge may cause an explosion
hazard.
Avoid any actions that cause the generation of
electrostatic charge, such as rubbing with a dry
cloth on coating face of product.
WARNING
Theinstrumentmodicationorpartsreplacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
willvoidthecertication.
Vista de pagina 181