12-7
IM 01E30D01-01EN
<12. OUTLINE>
Accuracy for Multi-drop of HART
(generally accepted values)
Vs: Span setting value (m/s)
Size in mm (in.)
Spaninm/s(ft/s)
Accuracy
25 to 200
(1 to 8)
0.3
≤Vs<1
(1
≤Vs<3.3)
±(0.4+0.3/Vs)% of span
1≤Vs<2
(3.3≤Vs<6.7)
± 0.5% of span
2≤Vs≤10
(6.7≤Vs≤33)
± 0.25% of span
(at indications less
than 50% of span)
± 0.5% of rate
(at indications 50% or
more of span)
Repeatability (Reference):
±0.2% of rate
(When
theowvelocityis1.5m/stoward2m/sof
setting span)
Insulation Resistance:
Inspection location Terminal
Test voltage Standard
Power Supply/Digital
Output - Functional
Grounding
SUPPLY/DO
500 V DC
100MΩ
or more
When the optional code A is selected (with the lightning protec-
tor), values are as follows.
Power Supply/Digital
Output - Functional
Grounding
SUPPLY/DO
100 V DC
20MΩ
or more
Note: Conduct the test according to the instruction manual.
Dielectric Strength:
Inspection location Terminal
Test
voltage
Test
time
Standard
Power Supply/Digital
Output - Functional
Grounding
SUPPLY/DO
500 V
AC
1
min.
25
mA
or less
When the optional code A is selected (with the lightning protec-
tor), values are as follows.
Power Supply/Digital
Output - Functional
Grounding
SUPPLY/DO
100 V
AC
1
min.
6 mA
or less
Note: Conduct the test according to the instruction manual.
Safety Requirement Standards:
EN61010-1, CAN/CSA C22.2 No. 61010-1-04
Altitude of installation site: Max. 2000 m above sea
level
Installation category: I
“Overvoltage category (Installation category)”
describes
anumberwhichdenesatransient
overvoltage condition. It implies the regulation for
impulse withstand voltage. “ I ” applies to electrical
equipment which is supplied from the circuit when
appropriate transient overvoltage control means
(interfaces) are provided.
Pollution degree: 2
“Pollution degree” describes the degree to which
a solid, liquid, or gas which deteriorates dielectric
strength or surface resistivity is adhering. 2applies
to normal indoor atmosphere. Normally, only non-
conductive pollution occurs. Occasionally, however,
temporary conductivity caused by condensation must
be expected.
Indoor/Outdoor use
EMC Conformity Standards:
,
EN61326-1 Class A, Table 2 (For use in industrial
locations)
EN61326-2-3
CAUTION
This instrument is a class A product, and it is de-
signed for use in the industrial environment.
Please use this instrument in the industrial environ-
ment only.
Pressure Equipment Directive:
Module:
H
Type of Equipment: Piping
Type of Fluid: Liquid and Gas
Group of Fluid: 1 and 2
General-purposeUse/ExplosionProofType;
MODEL
DN
(mm)
(*1)
PS
(MPa)
(*1)
PS · DN
(MPa ·
mm)
CATEGORY
(*2)
AXR025G/C
25 4 100
Article 3 (*3)
paragraph 3
AXR040G/C 40
4 160 II
AXR050G/C 50
4 200 II
AXR065G/C 65
2 130 II
AXR080G/C 80
2 160 II
AXR100G/C 100
2 200 II
AXR150G 150
2 300 II
AXR200G 200
2 400 III
*1: PS: Maximum allowable pressure for Flowtube
DN: Nominal size
*2: Fordetails,see“Table6CoveredbyANNEXIIofEC
Directive on Pressure Equipment Directive 97/23/EC.”
*3: AXR025G/CisoutsidethescopeofCEmarkingforPED.
SILCertication:
AXRseriesarecertiedbyTÜVincompliancewith
the following standards;
IEC 61508: 2000; Part1 to Part7
Functional Safety of Electrical/electronic/programma-
ble electronic safety-related systems; SIL 2 capabil-
ity
forsingleowmeteruse,SIL3capabilityfordual
owmeteruse.