<14. EXPLOSION PROTECTED TYPE INSTRUMENT>
14-7
IM 01E30D01-01EN
Intrinsically safety circuit
Maximum Voltage (Uo): 14 V
Maximum Current (Io): 17 mA
Maximum Electrical Power (Po): 0.12 W
Fluid Temperature: -20 to 130°C
Ambient Temperature: –20 to 55°C
Grounding:
Groundingresistanceof10Ωorlessis
necessary for class A grounding terminal.
Groundingresistanceof100Ωorlessis
necessary for Functional grounding terminal.
When the optional code A is selected,
groundingresistanceof10Ωorlessshallbe
required.
IMPORTANT
Class A grounding should be installed in non-
hazardrous area.
Flameproot packing adapter:
The specifying optional code G11 is necessary.
Incaseoftwoameproofpackingadapters,
select optional code G32 with G11.
(2) Wiring Installation
Fortheexternalwiringofameprooftypes,usea
ameproofpackingadapter(optionalcodeG11,
G32) approved by Yokogawa (refer to Figure
14.5.2) (refer to “INSTALLATION AND OPERATING
PRECAUTIONS FOR TIIS FLAMEPROOF
EQUIPMENT” at the end of this manual).
WARNING
Availableonlyforwiringofusingaameproof
packing adapter approved by Yokogawa.
TheameproofmetalconduitwiringforTIIS
ameprooftypeisnot permitted.
(2-1) Wiring Cable through Flameproof Packing
Adapter
WARNING
FortheTIISameprooftypewithwiringusinga
ameproofpackingadapter,wirecablesthrough
the packing adapters approved by Yokogawa
(optional code G11, G32).
T1
T2
L
18(0.71)
G
F
D
C
16.5(0.65)
*Packing
(Choose from the table below
depend on cable outside diameter)
Unit : mm
(Approx. inch)
Adapter body(M. Screw)
Packing case
Hexagon socket set screw
O-Ring
Packing *
O-Ring
O-Ring
Washer
Union nut
Packing gland
Clamp ring
Clamp nut
O-Ring
B.coupling
Cable(user's scope)
Hexagon socket set screw
0.26
(0.57)
ø8.0 to ø10.0 (0.31 to 0.39)
ø10.0 to ø12.0 (0.39 to 0.47)
16 8-10
16 10-12
Identification
mark
Weight
kg
(lb)
Packing diameter
Cable outer diameter
T1 T2 C D
L
Dimension
G 1/2 G 1/2
35
(1.38)
39
(1.54)
94.5
(3.72)
ø10.0(0.39)
ø12.0(0.47)
ø20.0
(0.79)
F G
F1405.ai
Figure 14.5.1 Flameproof Packing Adapter
Apply a nonhardening sealant to the terminal
box connection port and to the threads on the
ameproofpackingadapterforwaterproong.
Flameproof
packing adapter
Wiring metal
conduit
Flexible metal conduit
Tee
Drain plug
F1406.ai
Apply a
non-hardening
sealant to the threads
for waterproofing.
Figure 14.5.2 Typical Wiring Using Flexible Metal
Conduit
Followtheprocedureforameproofpacking
adapter setting. (refer to Figure 14.5.3)
CAUTION
Beforeghting,conrmcablelengthfrom
terminaltoameproofpackingadapterwhen
setting. Once it is tightened, loosening and re-
tightening may damage its sealing performance.
(a) Loosen the locking screw and remove the
terminal box cover.
(
b) Measure the cable outer diameter in two
directions to within 0.1 mm.
(c) Calculate the average of the two diameters, and
use packing with an internal diameter nearest to
this value (see Table of the Figure 14.5.1).
(d) Screwtheameproofpackingadapterintothe
terminal box until the O-ring touches the wiring
port(atleast6fullturns),andrmlytightenthe
lock nut.