5-17
IM WX104-01E
Configuring the DX100/DX200/MV100/MV200
1
2
3
4
5
6
7
8
9
Index
Scan Interval/Memory
Set these parameters when the data type is set to [EVENT & DISP] or [EVENT]
Click here (also selectable from [Setting] - [SETUP [Basic] Setting]-[Setting])
Check the channels you
wish to sample
A/D Integrate
100 ms can be selected only when the scan interval is set to 2 s.
Scan interval
The selectable scan intervals vary depending on the model as follows:
DX102, DX104, DX204, DX208, MV102, MV104, MV204, and MV208: 125 ms and 250
ms
DX106, DX112, DX210, DX220, DX230, MV106, MV112, MV220, and MV230: 1s and 2 s
Memory Sample (save method of measured/computed data)
Number of blocks
When the data type is [EVENT], select 1, 2, 4, 8, or 16.
When the data type is [EVENT&DISP], select 1, 2, or 4.
Pre-Trigger Length
If 0% is selected, the event le will entirely consist of data after the trigger. If 100% is
selected, the event le will entirely consist of data before the trigger.
Memory Sample
Select the channels that are to be saved to the memory.
Note
If [Save] is set to [Manual], the data directory is created at a location that cannot be managed
by the DAQ Desktop. Therefore, the DAQ Desktop cannot be used to handle data files in that
directory.
5.5 Configuring the Setup Mode